BeautyTech Manual

Whole Counter Mode

One of the most important functions in the new Beauty device with plasma is the whole counter mode. The market is the lack of digital electronic control of this device in terms of the duration of the arc. Duration of the arc is one of the most important points in the use of this type of devices for aesthetic purposes. Because the duration of the arc will have repercussions on the amount of energy you put into the skin and then if you actually place whatever arc or whatever frequency and whatever power label for long enough, you can actually go into the bone. For example, if you have got a very low power arcing device and you place the arc for long enough, for example, 20 seconds, you make your bets you are going into the bone.

So in our new device, you’re able to control to the milliseconds the amount of the arcing you are going to put onto the skin.

Basically the new function is called the whole counter, so what you are able to do is to precisely and digitally control the timing of the arc. Why is that important? This is important because at this time the arc is controlled either using a button or using your own hands and the counting mechanism is your eyes and you’re not able to count the millisecond and also your hands are not as precise either whereas a computer is and we have made this available in our new device.

You are now able to set the exact timing of the arcing and the energy will be delivered in digitally controlled system. The longer you are actually applying the arc the more energy you are applying on the skin and in the shortest time the less energy you are applying into the skin. you are unable to perfect measurement of arc whereas a computer does. This is the most accurate and safest energy delivering device in the market because you are not relying on your eyes, on your hands or on your judgment—instead you are relying on a computer for the duration of the arc .

How ARC Works?

  • The target time is set 600 milliseconds for arcing in whole counter mode. We want the exact duration of the arc to be 600 milliseconds and then to shut off.
  • You have the arc in the region of 200 milliseconds. As our hands are not very precise or you may move your hands unintentionally, then the arc will shut off and when the arc is off then the counter stops counting.

  • So, There are still 400 milliseconds to keep the arc on and again it is managed to get the arc on for 50 milliseconds because your hand is not very good and trembling, so the arc again switched off inadvertently for in the region of 10 milliseconds.

  • Because of repeatedly on and off, the computer counts 200 plus 50 plus 60 which is 310 milliseconds. As because of the timing is set for 600 milliseconds, the tip of the device will not shut off and wait for another 290 milliseconds of timing in order to shut off.

  • When you move your hand and you manage to get the arc on, if it is on then after remaining 290 milliseconds the device will automatically shut off.

Whole Counter Example & Explention in Video

Start Counter

The importance of the start counter is the same as the whole counter. Both are giving the option to control the duration of the arc digitally. Because the duration of the arc as you must know already is extremely important and the outcome of aesthetic procedures using arcane devices and plasma devices.

We have already read about the importance of the arc duration and plasma devices for aesthetic uses and digital control because the arc duration is proportional to the amount of energy that you put into the skin very very important and also it’s proportional to the intensity the treatment we know that the hand control the eye feedback look the counting the timing of arc using your brain and your eyes is not accurate a computer does a match better job.

The Main difference between the start counter and the whole counter is that with start counter we are controlling the maximum duration of the arc. We could also call it max counter or maximum time of arcing

Increase its safety – We control the maximum duration of the arc so its increase the safety so let’s take for an example if you are doing eyelid tightening and you want to make sure that the arc doesn’t much more than for example 600 milliseconds. You are never going to exceed the duration of 600 milliseconds.

You can set the maximum duration of the arc. So we can set it up from 30mm, 40mm so on after whatever length you want.

How start counter works?

  • let’s take for an example you are setting the start counter to count the maximum duration of the arc for 600 milliseconds.
  • What the device does Switch ON the needle and waits for the arc to take place it does not start counting up until it detects that the arc is take place when the arc is detected then it starts counting.
  • let’s assume we set the device at 600 milliseconds. So when the arc is ON so computer starts immediately counting keeps the tip of the device ON just as long as 600 milliseconds no more no less.
  • After the arc is detected it does not matter whether arc was ON for 10 milliseconds or 500 milliseconds or 590 milliseconds. it will keep the tip ON for 600 milliseconds no matter what.
  • That’s really very important to understand this thing.

Start Counter Example & Explention in Video

Difference between Whole Counter and Start Counter

Whole Counter

  • Sets the Maximum Arc Time

In the whole counter what we do is actually counting the total time of arcing for example 600 milliseconds the device will make sure that the arc duration in total 600 milliseconds it doesn’t matter the interval in between when arc is off it will still wait whether arc is off for 10, 20, 50 milliseconds till wait to arc to take place (ON) up until counts and overall 600 milliseconds.

The whole counter we call whole counter because the whole counter function comes whole of the time and make sure time you set the device to be operational the actually produce the arc is actually producing the arc for the exact time in total.

Start Counter

  • Set the Whole Time of the Arc

In start counter Soon as the arc, start to take place the counter starts and its finishes exactly at 600ms when the device switches Off the power to the tip-off. So start counter in simple words is the device starts counting when it detects the arc take place and it does not matter how much takes place for 5 milliseconds, 1m, 10mm, it will start counting at that point no matter what so this way we called start the counter.

So start counter gives you the option of making sure you have a limit to the time to the arcking time that is limit that is digitally controlled its not a limit that you can control using your eyes using your hands on your perception in general that as a human is not that good I mean so this why computer help us.

Power Levels

Which is variable frequency, first of all, we would like to run your through the basic circuit designed. In new Beautyteck device, we have some actually changed the design of the device we have a completely different concept of delivering the very best plasma device today’s so what we have done has just changed the transformer is the power. Right now we are using a very good step-up transformer. From the one side of transformer, there is a digital entrance input a sine wave which is digitally created. Therefore we can change that digitally its voltage and frequency.

The Output  we are going to get the same sine wave but a  higher voltage and at  the same frequency so what we end up with is  a device that’s got variable voltage and variable frequency and the beautyteck of it is that the variable frequency goes throughout a very vast range now this is not  trivial this is non-trivial at all  because  no other device that is on the market at the moment has even got a variable frequency and certainly no other device is going to have for a long time the same range of frequencies that we have  got.

Why Variable Frequency

The main purpose is because when you conveyed the frequency you can also vary the depth of ablation provided that the voltage is the same. So Higher the frequency the less degree of ablation the lower the frequency the higher is the degree of ablation keeping the voltage and the arc duration the same and this is a very important function of the new device.

 

What is the frequency ?

Many People actually confused the frequency with something else was the power level and some of the steps and many other things in order to make it clear to you what we get at the output of our plasma device and most of other plasma devices as well is a smiple sine wave and that is the case for almost all the plasma devices